how to measure free chlorine with the ultrapen pt4

how to measure free chlorine with the ultrapen pt4

OPERATING INSTRUCTIONS Ultrapen PT4 Free Chlorine Pen
NOTE: Selecting “ESC” from any menu immediately powers the PT4 off without saving changes.

I. Temperature Unit Selection
The PT4 allows you to select the type of units used for displaying temperature:
˚C (Degrees Celsius) or ˚F (Degrees Fahrenheit).
To set the preference:
1. Push and release the push button to turn the PT4 on.
2. Push and hold the button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL”, “PAr SEL”, “SOL ck”, and “ESC”.
3. Release the button while “ºCºF TEMP” is displayed. The display will alternate between “PUSHnHLD” and “ºCºF TEMP”.
4. Push and hold the button. The display will alternate between “˚C”, “˚F” and “ESC”.
Release the button when desired unit preference displays.
5. “SAVEd ºC” or “SAVEd ºF” will display; then the unit will power off.

II. FCE Mode Selection
The PT4 allows you to select the FCE measurement mode you prefer:
Hold Mode (default) — will display real-time readings until stable or 2 minutes, which ever
comes fi then display fi readings.
LIVE Mode — real-time readings are displayed continuously for up to 5 minutes, a push and release of the button will turn your PT4 off immediately.
To set the FCE measurement mode preference:
1. Push and release the push button to turn the PT4 on.
2. Push and hold the button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL”, “PAr SEL”, “SOL ck”, and “ESC”.
3. Release the button when “ModE SEL” is displayed. The display will alternate between “PUSHnHLD” and “ModE SEL”.
4. Push and hold the push button. The display will alternate between “Hold”, “LIVE” and “ESC”.
5. Release the button when desired mode displays.
6. “SAVEd” will display, then the PT4 will power off.

III. FCE Measurement
The following table explains what the LED Indicator Light signals indicate and gives the duration of each signal:

Capture

 

 

 

CAUTION: To measure solution at the extremes of the specified temperature or FCE range, allow the PT4 to equilibrate by submerging the sensor in the sample solution for 1 minute prior to taking a measurement.

NOTE: If you cannot dip the PT4 in the sample solution, pour the sample into a clean container. If you don’t have a container or need to test a vertical stream of solution, use the scoop to hold sample solution.
1. Rinse the pen 3 times in a sample of the solution.
2. Push and release the push button.
3. While the LED flashes rapidly, dip the PT4 in FRESH sample solution so that the sensor is completely submerged. If you do not submerge the sensor in solution before the flashing slows, allow the PT4 to power off and retake the reading.
4. While the LED flashes slowly, swirl the PT4 around to remove any air bubbles, keeping the sensor submerged.
a. In Hold mode when the LED turns on solid, remove the PT4 from solution. The display will alternate between the final FCE and temperature readings. Note the readings for your records.
b. In LIVE mode allow the PT4 to remain in solution while the LED flashes slowly. The display will alternate between live FCE and temperature readings. Note the readings for your records. LIVE measurement will time out after 5 minutes OR push and release the push button to turn the PT4 off at any time during LIVE measurement.