OPERATING INSTRUCTIONS Ultrapen PT3 ORP Pen
NOTE: Selecting “ESC” from any menu immediately turns the PT3 off without saving changes.

I. Temperature Unit Selection
The PT3 allows you to select the type of units used for displaying temperature:
°C (Degrees Celsius) or °F (Degrees Fahrenheit).
To set the preference:
1. Push and release the push button to turn the PT3 on.
2. Push and hold the button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
3. Release the button while “ºCºF TEMP” is displayed. The display will alternate between “PUSHnHLD” and “ºCºF TEMP”.
4. Push and hold the button. The display will alternate between “°C”, “°F” and “ESC”. Release the button when desired unit preference displays.
5. “SAVEd ºC” or “SAVEd ºF” will display; then the PT3 will turn off.

II. ORP Mode Selection
The PT3 allows you to select the ORP measurement mode you prefer:
Hold Mode (default) — once stabilized, the readings are captured then displayed.
LIVE Mode — real-time readings are displayed continuously during measurement.
To set the pH measurement mode preference:
1. Push and release the push button to turn the PT3 on.
2. Push and hold the button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
3. Release the button when “ModE SEL” is displayed. The display will alternate between “PUSHnHLD” and “ModE SEL”.
4. Push and hold the push button. The display will alternate between “Hold”, “LIVE” and “ESC”.
5. Release the button when desired mode displays.
6. “SAVEd” will display, then the PT3 will turn off.

III. ORP Measurement
The following table explains what the LED Indicator Light signals indicate and gives the duration of each signal:

Capture

 

 

 

 

CAUTION: To measure solution at the extremes of the specified temperature or ORP range, or when changing between solutions that have extreme differences in ORP values, or from high to low ionic strength: Rinse the sensor thoroughly with clean water then allow it to soak in a sample of the solution to be tested for a minimum of 1 minute.
(In some applications you may want to allow the sensor to soak in solution for an hour or more.)
NOTE: If you cannot dip the sensor in the sample solution, pour the sample into a clean container. If you don’t have a container or need to test a vertical stream of solution, use the scoop to hold sample solution.

1. Rinse the PT3. If measuring from a container, submerge the sensor and swirl it around in FRESH sample solution 3 times. Alternatively, 30 seconds under a stream or swirling in a body of water is sufficient to prepare the sensor for measurement.
2. Remove PT3 from solution. (Fill the container one more time with FRESH sample solution, if applicable.) Then push and release the push button.
3. Grasp the PT3 by its case with your fingers positioned between the display
and the battery cap to avoid sample contamination.
4. While the LED flashes rapidly, dip the PT3 in FRESH sample solution so that
the sensor is completely submerged.
5. While the LED flashes slowly, swirl the PT3 around to remove bubbles,
keeping the sensor submerged.
a. In Hold mode when the LED turns on solid, remove the PT3 from solution. The display will alternate between the final ORP and temperature readings. Note the readings for your records.
b. In LIVE mode allow the PT3 to remain in solution while the LED flashes slowly. The display will alternate between live ORP and temperature readings. Note the readings for your records. LIVE measurement will time out after 90 seconds OR push and release the push button to turn the PT3 off at any time during LIVE measurement.