Ultrapen PT1 Conductivity, TDS and Salinity Calibration: Myron L Meters

Posted by 3 Apr, 2014

TweetThe Ultrapen PT1 is designed to be very reliable and requires only infrequent calibration. The factory recommends calibrating each measurement mode you use once monthly. However, you should check the calibration whenever measurements are not as expected. The PT1 is programmed for 2 calibration options: Wet Calibration or Factory Calibration. Wet calibration is most accurate. […]

The Ultrapen PT1 is designed to be very reliable and requires only infrequent calibration. The factory recommends calibrating each measurement mode you use once monthly. However, you should check the calibration whenever measurements are not as expected. The PT1 is programmed for 2 calibration options: Wet Calibration or Factory Calibration. Wet calibration is most accurate. But if a high quality standard KCl-1800 µS or 442-3000 ppm solution is not available, the PT1 can be returned to factory settings.

A. Wet Calibration
Use calibration solution specified for measurement mode: Use KCL- 1800 for Cond KCl; Use 442-3000 for tdS 442, SALt 442, tdS NaCl, and SALt NaCl. See Specifications table for 442 solution ppm NaCl equivalent value. Calibrating TDS simultaneously calibrates SALT for the same value and vice versa.
1. Pour calibration solution into a clean container.
2. Rinse the pen 3 times by submerging the cell in fresh calibration solution and swirling it around.
3. Remove pen from solution, then fill the container one more time.
4. Press and release the push button. The LCD will briefly display the firmware version then the current measurement mode. Ensure the PT1 is in the correct solution mode.
5. Immediately push and hold the push button. The display will scroll through “CAL”, “SOL SEL”, “FAC CAL”, “ºCºF TEMP”, and “ESC”. Release the button when “CAL” displays.
6. Grasp the pen by its case with your fingers positioned between the
display and the pen cap to avoid sample contamination.
7. While the LED flashes rapidly, dip the pen in calibration solution so that the cell is completely submerged. If you do not submerge the cell in solution before the fl slows, allow the pen to power off and start over.
8. While the LED flashes slowly, swirl the pen around to remove bubbles, keeping the cell submerged. Keep pen at least 1 inch (2½ cm) away from sides/bottom of container.
9. When the LED light stays on solid, remove the pen from the solution. “CAL SAVED” will display indicating a successful calibration.
Note: If an incorrect solution is used or the measurement is NOT within calibration limits for any other reason, “Error” displays alternately with “CLEAn CEL/CHEC SOL”. Check to make sure you are using the correct calibration solution. If the solution is correct, clean the cell by submerging the cell in a 1:1 solution of Lime-A-Way® and water for 5 minutes. Rinse the cell and start over.
10. Small bubbles trapped in the cell can give a false calibration. Measure the calibration solution again to verify correct calibration. If the reading is not within ±1% of the calibration solution value, repeat calibration.

B. Factory Calibration
If you do not have the proper calibration solution or wish to restore the pen to its original factory settings for any other reason, use the FAC CAL function to calibrate the PT1.
1. Press and release the push button. The LCD will briefly display the firmware version then the current measurement mode.
2. Immediately push and hold the push button. The display will scroll through “CAL”, “SOL SEL”, “FAC CAL”, “ºCºF TEMP”, and “ESC”. Release the button when “FAC CAL” displays.
3. While the display scrolls through “PUSHnHLD” and “FAC CAL”, push and hold the push button until the display scrolls through “SAVEd” and “FAC CAL”, indicating the pen has been reset to its factory calibration.
4. Allow the pen to time out to turn power off.

MyronLMeters.com is the premier internet retailer of the Ultrapen PT1 and other reliable Myron L meters. Save 10% on Myron L meters when you order online HERE.

Categories : Product Updates

Ultrapen PT1 Conductivity, TDS and Salinity Pen Measurement: MyronLMeters.com

Posted by 3 Apr, 2014

TweetI. solution selection The PT1 allows you to select from several preprogrammed measurement modes. The following table lists measurement modes with their corresponding parameters; temperature compensation and TDS conversion solution models; and units of measure. Mode Parameter solution Model units Cond KCl Conductivity potassium chloride microsiemens (µS) tds 442 Total Dissolved Solids (TDS) 442™ Myron L […]

I. solution selection

The PT1 allows you to select from several preprogrammed measurement modes. The following table lists measurement modes with their corresponding parameters; temperature compensation and TDS conversion solution models; and units of measure.

Mode Parameter solution Model units
Cond KCl Conductivity potassium chloride microsiemens (µS)
tds 442 Total Dissolved Solids (TDS) 442™ Myron L NaturalWater Standard parts per million (ppm)
tds NaCl TDS sodium chloride ppm
salt 442 Salinity 442™ Myron L NaturalWater Standard parts per thousand (ppt)
salt NaCl Salinity sodium chloride ppt

 

esc This is the escape function. Selecting escape exits solution selection without saving changes and turns the PT1 off.

 

To select a measurement mode:
1. Press and release the push button. The LCD will briefly display the firmware version then the current measurement mode. If the measurement parameter and solution type displayed are correct, proceed to Temperature Unit Selection.

If not, proceed to step 2.

2. Immediately push and hold the push button. The display will scroll through “CAL”, “SOL SEL”, “FAC CAL”, “ºCºF TEMP”, and “ESC”. Release the button when “SOL SEL” displays.

3. While the display scrolls through “PUSHnHLD” and “SOL SEL”, push and hold the push button. The display will scroll through “Cond KCl”, “tdS 442”, “tdS NaCl”, “SALt 442”, “SALt NaCl” and “ESC”. Release when the desired measurement mode displays.

4. “SAVED” displays indicating the measurement mode is saved in memory. Allow the pen to time out to turn power off.

II. Temperature Unit Selection

The PT1 allows you to select the type of units used for temperature measurements. The following table lists preference options with their corresponding units.

Mode Unit Preference

C Degrees Celsius

F Degrees Fahrenheit

esc This is the escape function. Selecting escape exits temperature unit selection without saving changes and turns the PT1 off.

To set the preference:
1. Press and release the push button. The LCD will briefly display the firmware version then the current measurement mode.
2. Immediately push and hold the push button. The display will scroll through “CAL”, “SOL SEL”, “FAC CAL”, “ºCºF TEMP”, and “ESC”. Release the button when “ºCºF TEMP” displays.
3. While the display scrolls through “PUSHnHLD” and “ºCºF TEMP”, push and hold the push button. The display will scroll through “C”, “F” and “ESC”. Release when the desired unit option displays.
4. “SAVED” displays indicating the unit preference is saved in memory.
Allow the pen to time out to turn power off.

III. Normal Operation

Before you take a reading, make sure the pen is clean, calibrated and in the appropriate measurement mode. The sample solution must also be within the specified measurement range. Keep all foreign material away from the sample to avoid contamination.

Note: If you cannot dip the pen in the sample solution, pour the sample into a clean container. If you don’t have a sample container and need to test a vertical stream of solution, use the scoop.

The following table explains what the LED Indicator Light signals mean and gives the duration of each signal.

LED Indicator Light Signal Meaning duration

Rapid Flashing Dip pen in solution 6 sec

Slow Flashing Measurement in process 10-20 sec

Solid Light Note measurement value 6 sec

CAUTION: To measure solution at the extremes of the specified temperature range, allow the pen to equilibrate by submerging the cell in the sample solution for 1 minute prior to taking a measurement.

1. Rinse the pen 3 times by submerging the cell in fresh sample solution and swirling it around.

2. Remove pen from solution, then press and release the push button. Firmware version will be displayed, then current measurement mode.

3. Grasp the pen by its case with your fingers positioned between the display and the pen cap to avoid sample contamination.

4. While the LED flashes rapidly, dip the pen in fresh sample solution so that the cell is completely submerged. If you do not submerge the cell in solution before the flashing slows, allow the pen to power off and retake the reading.

5. While the LED flashes slowly, swirl the pen around to remove bubbles, keeping the cell submerged. Keep the pen at least 1 inch (2½ cm) away from sides/bottom of container, if applicable.

6. When the LED turns on solid, remove the pen from solution. The display will alternate between the measurement and temperature readings. Note the readings for your records.

MyronLMeters.com is the premier internet retailer of the Ultrapen PT1 and other reliable Myron L meters. Save 10% on Myron L meters when you order online HERE.

Categories : Product Updates

Ultrapen PT1 Conductivity, TDS, Salinity Pen Features: MyronLMeters.com

Posted by 3 Apr, 2014

TweetThe ULTRAPEN™ PT1 Conductivity/ TDS/Salinity Pen is designed to be extremely accurate, fast and simple to use in diverse water quality applications. Advanced features include the ability to select from 3 different solution types that model the characteristics of the most commonly encountered types of water; proprietary temperature compensation and TDS conversion algorithms; highly stable […]

The ULTRAPEN™ PT1 Conductivity/ TDS/Salinity Pen is designed to be extremely accurate, fast and simple to use in diverse water quality applications. Advanced features include the ability to select from 3 different solution types that model the characteristics of the most commonly encountered types of water; proprietary temperature compensation and TDS conversion algorithms; highly stable microprocessor-based circuitry; user-intuitive design; and waterproof housing. A true one-handed instrument, the PT1 is easy to calibrate and easy to use. To take a measurement, you simply press a button then dip the pen in solution. Results display in seconds.

Features
1. Push Button — turns instrument on; selects mode and unit preferences.
2. Pen Cap — provides access to battery for replacement.
3. Clip — holds pen to shirt pocket for secure storage.
4. Battery Indicator — indicates charge left in battery.
5. display — displays measurements, mode options and battery indicator.
6. LED Indicator Light — indicates when to dip instrument in solution, when measurement is in progress, and when to remove instrument from solution.
7. electrodes — measure electric current of solution.
8. Cell — contains flux field in defined area for accurate current
measurement.
9. scoop — contains sample solution for measurement when sampling from a vertical stream. To use, slide the open end of the scoop over the bottom of the pen until the neck of the scoop is flush with the top of the cell. Hold pen with scoop end under stream. Rinse and fill with sample solution 3 times. Fill with solution again, then take measurement. We recommend you recalibrate the pen using the scoop to retain accuracy of ±1%.

Technical Specs

Measurement Range: 1 – 9999 µS or ppm (0.0010 – 9.999 ppt salinity)
Accuracy (After Wet Calibration): ± 1% of reading
Repeatability: < 1000 µS or ppm ± 1 Count
≥ 1000 µS or ppm ± 0.3% of reading
Resolution: Conductivity and TDS:
0.1 for 1.0 – 99.9 µS or ppm
1 for 100 – 9999 µS or ppm
Salinity: 0.0001 for 0.0010 – 0.0999 ppt
0.001 for 0.100 – 9.999 ppt
Temperature: 0.1 ºC or ºF
Time to Reading Stabilization: 10 – 20 seconds
Active Mode Power Consumption: 30 – 100 mA
Sleep Mode Power Consumption: 2 µA
Temperature Measurement Range: 0 – 71° C or 32 – 160° F
Temperature Accuracy Displayed: ± 0.1 ºC or ± 0.1 ºF
Temperature Compensation Method: Automatic to 25ºC
Physical Dimensions: 17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D
Weight: 55 g or 1.94 oz.
Case Material: Anodized Aluminum with Protective Coating
Battery Type: N type, Alkaline
Battery Voltage: 1.5 V
Calibration Solution Point: 1800 µS KCl; 3000 ppm 442™ (2027 ppm NaCl)
Operating/Storage Temperature: 0 – 55ºC or 32 – 131ºF
Water Resistance: IP67 and NEMA 6

Electrostatic discharge to case of instrument may cause PT1 to spontaneously power on. In this case, the PT1 will power off after several seconds.

MyronLMeters.com is the premier internet retailer of the Ultrapen PT1 and other reliable Myron L meters. Save 10% on Myron L meters when you order online HERE.

Categories : Product Updates

End of Year Sale

Posted by 4 Dec, 2013

Tweethttp://us2.campaign-archive1.com/?u=8c3e1105d354c0c814b53a0e0&id=f52da0f404&e=UNIQID

http://us2.campaign-archive1.com/?u=8c3e1105d354c0c814b53a0e0&id=f52da0f404&e=UNIQID

Categories : Deal of the Month, MyronLMeters.com Valued Customers

Myron L Meters for Hydroponics: MyronLMeters.com

Posted by 5 Sep, 2013

TweetFeatures • Handheld meters measure TDS and/or pH • Monitor measures TDS • All instruments are easy to operate and calibrate • High degree of accuracy • Immediate results • Kit comes with solutions required to calibrate • Temperature compensated readings TDS Monitoring The nutrient solution and its management are the foundation of a successful […]

Features

• Handheld meters measure TDS and/or pH
• Monitor measures TDS
• All instruments are easy to operate and calibrate
• High degree of accuracy
• Immediate results
• Kit comes with solutions required to calibrate
• Temperature compensated readings

TDS Monitoring

The nutrient solution and its management are the foundation of a successful hydroponics system. The function of a hydroponics nutrient solution is to supply the plant roots with water, oxygen and essential mineral elements in soluble form.

A test of the Total Dissolved Solids (TDS) using the DS Meter or pDS Meter or continuous monitoring with the HYDRO-STIK gives the grower accurate measurements of the concentration of nutrients in solution. If the concentration drops below the optimum level required to sustain and grow the plants, add more nutrient- rich solution until the desired concentration level is achieved. This prevents haphazard dosing and wasted solution, which minimizes costs to the grower.

pH Monitoring

pH of the nutrient solution is also critical to successful plant growth. All elements have a specific solubility pH range. This means that mineral elements dissolve and can become more concentrated in solution within certain pH ranges. Roots absorb only the dissolved nutrients, so this is critical to plant growth.
The TH1H and the pDS Meter quickly and easily measure pH.

Monitoring the addition of a pH balancing solution with the proper meter lets the grower precisely adjust the pH level.

Beyond affecting nutrient availability, extremely low or high pH can even damage or kill plants.

All Myron L TDS and pH meters give lab-accurate results in the field.

All Myron L meters use advanced Temperature Compensation (TC) circuitry and equations to give you the best TC correction available.

Ultrapen PT2 pH and Temperature Pen

Ultrapen PT2 pH and Temperature Pen

Ultrapen PT1 TDS Pen

Ultrapen PT1 TDS Pen

T6/pH TDS and pH Meter

T6/pH TDS and pH Meter

Techpro II - TPH1 TDS, pH, Conductivity, Temperature

Techpro II – TPH1 TDS, pH, Conductivity, Temperature

PSTK Soil Test Kit

PSTK Soil Test Kit

 

 

Categories : Application Advice

The Ultrapen PT-3: MyronLMeters.com

Posted by 12 Jun, 2013

Tweet Myron l Meters Ultrapen PT-3 from Myron L Meters

Myron l Meters Ultrapen PT-3 from Myron L Meters
Categories : Product Updates

The Ultrapen PT-2: MyronLMeters.com

Posted by 12 Jun, 2013

Tweet Myron l Meters Ultrapen PT-2 from Myron L Meters

Myron l Meters Ultrapen PT-2 from Myron L Meters
Categories : Product Updates

New MyronLMeters.com SlideShare: The Ultrapen PT-1

Posted by 11 Jun, 2013

Tweet Myron l Meters Ultrapen PT-1 from Myron L Meters

Myron l Meters Ultrapen PT-1 from Myron L Meters
Categories : Product Updates

Water Quality Testing in RO Systems – MyronLMeters.com

Posted by 10 May, 2013

Tweet Water quality testing is vital to the design of an efficient, cost-effective RO system, and is one of the best ways to preserve system life and performance. Using an accurate Total Dissolved Solids (TDS) measurement to assess the system load prevents costly mistakes up front. The TDS measurement gives users the information they need […]

DH-UMIII-9PTK-2T

Water quality testing is vital to the design of an efficient, cost-effective RO system, and is one of the best ways to preserve system life and performance.

Using an accurate Total Dissolved Solids (TDS) measurement to assess the system load prevents costly mistakes up front. The TDS measurement gives users the information they need to determine whether or not pretreatment is required and the type of membrane/s to select. Ultrameter™ and ULTRAPEN PT1™ Series TDS instruments feature the unique ability to select from 3 industry standard solution models: 442 Natural Water™ NaCl; and KCl. Choosing the model that most closely matches the characteristics of source water yields measurements accurate enough to check and calibrate TDS monitor/controllers that can help alert to system failures, reducing downtime and increasing productivity. The same instruments provide a fast and accurate test for permeate TDS quality control. Measuring concentrate values and analyzing quality trends lets users accurately determine membrane usage according to the manufacturer’s specifications so they can budget consumption correctly. These daily measurements are invaluable in detecting problems with system performance where changes in the ionic concentration of post-filtration streams can indicate scaling or fouling. System maintenance is generally indicated if there is either a 10-15% drop in performance or permeate quality as measured by TDS.

Thin-film composite membranes degrade when exposed to chlorine. In systems where chlorine is used for microbiological control, the chlorine is usually removed by carbon adsorption or sodium bisulfite addition before membrane filtration. The presence of any chlorine in such systems will at best reduce the life of the membrane, thus, a target of 0 ppm free chlorine in the feedwater is desirable.

ORP gives the operator the total picture of all chemicals in solution that have oxidizing or reducing potential including chlorine, bromine, chloramines, chlorine dioxide, peracetic acid, iodine, ozone, etc. However, ORP can be used to monitor and control free chlorine in systems where chlorine is the only sanitizer used. ORP over +300 mV is generally considered undesirable for membranes. Check manufacturer’s specifications for tolerable ORP levels.

An inline ORP monitor/controller placed ahead of the RO unit to automatically monitor for trends and breakthroughs coupled with spot checks by a portable instrument will prevent equipment damage and failure. Myron L 720 Series II™ ORP monitor/controllers can be configured with bleed and feed switches as well as visible and audible alarms.

Ultrameter and ULTRAPEN portable handhelds are designed for fast field testing and are accurate enough to calibrate monitor/controllers. Our measurement methods are objective and have superior accuracy and convenience when compared to colorimetric methods where determination of equivalence points is subjective and can be skewed by colored or turbid solutions.

Monitoring pH of the source water will allow users to make adjustments that optimize the performance of antiscalants, corrosion inhibitors and anti-foulants. Using a 720 II Series Monitor/controller to maintain pH along with an Ultrameter Series or ULTRAPEN PT2™ handheld to spot check pH values will reduce consumption of costly chemicals and ensure their efficacy.

Most antiscalants used in chemical system maintenance specify a Langelier Saturation Index maximum value. Some chemical manufacturers and control systems develop their own proprietary methods for determining a saturation index based on solubility constants in a defined system. However, LSI is still used as the predominant scaling indicator because calcium carbonate is present in most water. Using a portable Ultrameter III 9PTKA™ provides a simple method for determining LSI to ensure the chemical matches the application.

The Ultrameter III 9PTKA computes LSI from independent titrations of alkalinity and hardness along with electrometric measurements of pH and temperature. Using the 9PTKA LSI calculator, alterations to the water chemistry can be determined to achieve the desired LSI. Usually, pH is the most practical adjustment. If above 7, acid additions are made to achieve the pH value in the target LSI. Injections are made well ahead of the RO unit to ensure proper mixing and avoid pH hotspots. A Myron L 720 Series II pH Monitor/controller will automatically detect and divert solution with pH outside the range of tolerance for the RO unit. ULTRAPEN PT2, TechPro II and Ultrameter Series instruments can be used to spot check and calibrate the monitor/controller as part of routine maintenance and to ensure uniform mixing.

Water hardness values indicate whether or not ion exchange beds are required in pretreatment. Checking hardness values directly after the softening process with the Ultrameter III 9PTKA ensures proper functioning and anticipates the regeneration schedule.

Alkalinity is not only important in its effect on the scaling tendency of solution, but on pH maintenance. Additions of lime are used to buffer pH during acid injection. Use a 9PTKA to measure alkalinity values for fast field analysis where other instrumentation is too cumbersome to be practical.

Though testing and monitoring pressure is a good way to evaluate system requirements and performance over time, measuring other water quality parameters can help pinpoint problems when troubleshooting. For example, if the pressure differential increases over the second stage, the most likely cause is scaling by insoluble salts. This means that any degradation in performance is likely due to the dissolved solids in the feed. Using a 9PTKA to evaluate LSI and calculate parameter adjustments is a simple way to troubleshoot a costly problem.

Myron L Meters saves you 10% on all Ultrameters and Ultrapens when you order online at MyronLMeters.com, where you can find the complete selection of Myron L meters, including the Ultrameter III 9PTKA.

Original story from International Filtration News V 32, no. 2

 

Categories : Application Advice, Case Studies & Application Stories, Technical Tips

Pool Draining Tips to Protect Water Quality – MyronLMeters.com

Posted by 1 May, 2013

TweetMyronLMeters.com has the most advanced lineup of pool analysis meters for the professional pool maintenance technician from the Ultrapen to the PoolPro PS9. Pool Draining Tips to Protect Water Quality With summer right around the corner, many swimming pool owners will be readying their swimming pools in anticipation of the season’s heat. As part of […]

MyronLMeters.com has the most advanced lineup of pool analysis meters for the professional pool maintenance technician from the Ultrapen to the PoolPro PS9.

Pool Draining Tips to Protect Water Quality

With summer right around the corner, many swimming pool owners will be readying their swimming pools in anticipation of the season’s heat. As part of this process, some pool owners like to drain old swimming pool water which has been sitting all winter. Though not a necessary task, the following tips are provided for you to properly drain pool water in order to protect the water resources in your community.

Whenever possible, it is best to drain your pool onto your landscape. This recycles your pool water, conserves irrigation water, and avoids the environmental risks associated with draining your pool to the street. Before draining your pool water to the street or to your landscape, be sure to follow the guidelines outlined below.

While draining pool water to the street is a common practice, it can prove harmful to the environment if the pool owner does not properly plan and prepare prior to draining. When pool water is drained to the street, it can carry other pollutants such as oil, grease, sediment, bacteria and trash down the storm drain and into the nearest creek, river, or the ocean. Swimming pool water also often contains harmful additives and chemicals. If the water is not properly treated to remove these pollutants prior to draining, they can cause further damage to the health of our waterways and to the plants and animals that live there.

Also, prior to draining to the street, residents are asked to sweep the curb and gutter between the discharge point from their yard to the storm drain down hill from their home. This will remove any pollutants from the gutter that may be carried up by the drained pool water to the storm drain.

For chlorine pools, chlorine levels must be lowered to less than 1 part per million prior to draining. This can be done naturally, by simply allowing the pool water to sit in the sun for a minimum of three days. Alternatively, de-chlorination kits can be purchased at home supply stores at a very reasonable cost. These kits have the tools you need to reach the appropriate chlorine levels before draining your pool to the curb and gutter.

Some people have salt water pool systems which may be preferred due to the lower amount of chemicals required for their operation. However, these pools must not be drained to the storm drain system due to their high salt content relative to the fresh water systems they drain into. Total dissolved solids (TDS) must be below 500 parts per million in order to drain into the street.

Green pools,” which are pools in which algae is growing, also must not be drained to the street. In these instances, algae must first be killed and removed. This is usually done by chlorinating the swimming pool until the algae is gone, then lowering chlorine to the allowable discharge level. Cartridge filters or diatomaceous earth (DE) filters should be rinsed onto a pervious surface such as landscaped areas or grass. While DE is actually beneficial in your garden, it can build up in storm drains and clog them. DE residues can be scooped up and simply thrown in the trash or put to use fending off worms in your garden.

For more information on how to reach acceptable chemical and TDS levels, call your pool maintenance specialist.

If you are a pool maintenance specialist, consider the PoolPro PS9TKA from MyronLMeters.com – the most advanced and comprehensive pool water analysis meter on the market.

PS9TK from MyronLMeters.com

 

 

 

 

 

 

 

 

 

Pool Pro PS9TK

Measures 9 Parameters: Conductivity, Mineral/Salts, TDS, Alkalinity, Hardness, LSI, pH, ORP/Free Chlorine, Temperature
LSI Calculator for hypothetical water balance calculations
Wireless data transfer capability with bluDock option
Auto-ranging delivers increased resolution across diverse applications
Adjustable Temperature Compensation and Cond/TDS conversion ratios for user-defined solutions
Nonvolatile memory of up to 100 readings for stored data protection
Date & time stamp makes record-keeping easy
pH calibration prompts alert you when maintenance is required
Auto-off minimizes energy consumption
Low battery indicator
(Includes instrument with case and solutions)