Conductivity Calibration on the Ultrameter II 6PIIFCe:

Posted by 23 Mar, 2014


*Note: This procedure applies to the Ultrameter, PoolPro, TechPro, and D-6 Dialysate meter.

a.  Fill and rinse the conductivity cell three times with a KCL standard solution. In this example, we’re using KCL-7000.

b. Refill conductivity cell with same standard solution you rinsed with.

c. Press






then press

CAL key





The “CAL” icon will appear on the display.













d. Press











to step the displayed value toward the standard’s value.

In this example, we’re pressing






to go down from 7032 to 7000. You can also hold a key down to scroll rapidly.

e. Press

CAL key






once to confirm the new value and end the calibration.