Conductivity Calibration on the Ultrameter II 6PIIFCe: MyronLMeters.com

Posted by 23 Mar, 2014

 

*Note: This procedure applies to the Ultrameter, PoolPro, TechPro, and D-6 Dialysate meter.

a.  Fill and rinse the conductivity cell three times with a KCL standard solution. In this example, we’re using KCL-7000.

b. Refill conductivity cell with same standard solution you rinsed with.

c. Press

COND

 

 

 

 

then press

CAL key

.

 

 

 

The “CAL” icon will appear on the display.

display

 

 

 

 

 

 

 

 

 

 

 

d. Press

Up

 

 

 

or

Down

 

 

 

 

to step the displayed value toward the standard’s value.

In this example, we’re pressing

Down

 

 

 

 

to go down from 7032 to 7000. You can also hold a key down to scroll rapidly.

e. Press

CAL key

 

 

 

 

 

once to confirm the new value and end the calibration.