I.      Calibration

The factory recommends calibrating the Ultrapen PT2 pH tester twice a month, depending on usage. However, you should check the calibration whenever measurements are not as expected. 3-point Wet Calibration is most accurate and is recommended.

NOTE: If the measurement is NOT within calibration limits for any reason, “Error” will display. Check to make sure you are using a proper pH buffer solution. If the solution is correct, clean the glass bulb of the sensor with a cotton swab soaked in isopropyl alcohol. Restart calibration.

NOTE: Small bubbles trapped in the sensor may give a false calibration. After calibration is completed, measure the pH buffer solutions again to verify correct calibration.

NOTE: If at any point during calibration, you do not submerge the sensor in solution

before the flashing slows, allow the pen to power off and start over.

A.    Calibration Preparation

  1. For maximum accuracy, fill 2 clean containers with each pH buffer. Arrange them in such a way that you can clearly remember which is the rinse solution and which is the calibration buffer. If you don’t have enough buffer, you can use 1 container of each buffer for calibration and 1 container of clean water for all rinsing. Always rinse the pH sensor between buffer solutions.

2.   Ensure the pH sensor is clean and free of debris.

B.    3-Point Calibration

Use pH 7, 4 and 10 buffers for 3-point calibration.

1. Thoroughly rinse the pen by submerging the sensor in pH 7 buffer rinse solution and swirling it around.

2. Push and release the push button to turn the unit on.
3. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
4. Release the button when “CAL” displays. The display will indicate “CAL” and the LED will flash rapidly. 
5. While the LED flashes rapidly, dip the pen in pH 7 buffer calibration solution so that the sensor is completely submerged. 
6. While the LED flashes slowly, the pH calibration point will display along with “CAL”. Swirl the pen around to remove bubbles, keeping the sensor submerged.
7.   If the pH 7 calibration is successful, the display will indicate “SAVED”, then “PUSHCONT” will be displayed.

8.   Push and release the push button to continue. The LED will begin flash rapidly.Repeat steps 5 through 8 with pH 4 and 10 buffer calibration solutions.After the 3rd calibration point is successfully saved, the display will indicate “SAVED” and power off.Verify calibration by retesting the calibration solution.

C.    2-Point Calibration

Use pH 7 and 4 or 10 buffers for 2-point calibration.

  1. Thoroughly rinse the pen by submerging the sensor in pH 7 buffer rinse solution and swirling it around.
  2. Push and release the push button to turn the unit on.
  3. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
  4. Release the button when “CAL” displays. The display will indicate “CAL” and the LED will flash rapidly.
  5. While the LED flashes rapidly, dip the pen in pH 7 buffer calibration solution so that the sensor is completely submerged.
  6. While the LED flashes slowly, the pH calibration point will display along with “CAL”. Swirl the pen around to remove bubbles, keeping the sensor submerged.
  7. If the pH 7 calibration is successful, the display will indicate “SAVEd”, then “PUSHCONT” will be displayed.
  8. Push and release the push button to continue. The LED will begin flashimg rapidly.

Repeat steps 5 through 7 with pH 4 or 10 buffer calibration solution.

Leave the pen in the same buffer solution until the unit powers off. The offset will be applied to the remaining calibration point.
Verify calibration by retesting the calibration solution.

 

D.    1-Point Calibration

Use pH 7, 4 or 10 buffer for 1-point calibration.

  1. Thoroughly rinse the pen by submerging the sensor in pH buffer rinse solution and swirling it around.
  2. Push and release the push button to turn the unit on.
  3. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
  4. Release the button when “CAL” displays. The display will indicate “CAL” and the LED will flash rapidly.
  5. While the LED flashes rapidly, dip the pen in pH buffer calibration solution so that the sensor is completely submerged.
  6. While the LED flashes slowly, the pH calibration point will display along with “CAL”; swirl the pen around to remove bubbles, keeping the sensor submerged.

If the pH calibration is successful, the display will indicate “SAVEd”, then “PUSHCONT” will be displayed. “PUSHCONT” will not display if you calibrated 4 or 10.

Leave the pen in the same buffer solution until the unit powers off. The offset will be applied to the remaining calibration points.

Verify calibration by retesting the calibration solution.

E.    Factory Calibration

When pH buffers are not available, the PT2 can be returned to factory default calibration using the FAC CAL function. This will erase any stored wet calibration. NOTE: default factory calibration resets the electronics only and does NOT take the condition of the sensor into consideration.

To  return your unit to factory calibration:

  1. Push and release the push button.
  2. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
  3. Release the button when “FAC CAL” displays. The display will alternate between “PUSHnHLD” and “FAC CAL”.
  4. Push and hold the push button. “SAVED FAC” displays indicating the pen has been reset to its factory calibration.