Search Results

Ultrapen PT4 Free Chlorine Pen Calibration: MyronLMeters.com

Posted by 4 Apr, 2014
how to calibrate free chlorine for the ultrapen pt4

how to calibrate free chlorine for the ultrapen pt4

IV. Calibration of the Ultrapen PT4 Free Chlorine Pen

The manufacturer recommends calibrating twice a month, depending on usage.

However, you should check the calibration whenever measurements are not as expected. For greatest accuracy, you should perform a 3-point wet pH calibration, and wet ORP calibration with the ORP Standard Solution closest in value to the solution you will be testing.

NOTE: If the measurement is NOT within calibration limits for any reason, “Error” will display. Check to make sure you are using a proper Myron L Company pH Buffer or ORP Standard Solution. If the solution is correct, clean the sensor as described in Sensor Cleaning section on page 4 of the operations manual. Restart calibration.

NOTE: Small bubbles trapped in the sensor may give a false calibration. After calibration is completed, measure the pH Buffer or ORP Standard Solutions again in solution check mode “SOL ck” (see pages 3 and 4 of the operations manual) to verify correct calibration.

NOTE: If at any point during calibration, you do not submerge the sensor in solution before the flashing slows, allow the PT4 to power off and start over.
NOTE: You should always calibrate with pH 7 first.

A. Calibration preparation
For maximum accuracy, fill 2 clean containers with each pH Buffer and/or ORP Standard Solution. Arrange them in such a way that you can clearly remember which is the rinse solution and which is the calibration standard/buffer. If you don’t have enough standard/ buffer, you can use 1 container of each standard/buffer for calibration and 1 container of clean water for all rinsing. Always rinse the FCE sensor between standard/buffer solutions. Ensure the FCE sensor is clean and free of debris.

B. pH Calibration using pH 7, 4, and 10 Buffer Solutions.
NOTE: You should always calibrate with pH 7 first.
1. Thoroughly rinse the PT4 by submerging the sensor in pH 7 Buffer rinse solution and swirling it around.
2. Push and release the push button to turn the PT4 on.
3. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL”, “PAr SEL”, “SOL ck”, and “ESC”.
4. Release the button when “CAL” displays.
5. The display will alternate between “PUSHnHLD” and “CAL.
6. Push and hold the button, The display will alternate between “PH” and “ORP”.
7. Release the button when “PH” is displayed.
8. The display will indicate “CAL” and the LED will flash rapidly.
9. While the LED flashes rapidly, dip the PT4 in pH 7 Buffer Calibration Solution so that the sensor is completely submerged.
10. While the LED flashes slowly, the pH calibration point will display along with “CAL”.
Swirl the PT4 around to remove bubbles, keeping the sensor submerged.
11. If the pH 7 calibration is successful, the display will indicate “SAVEd”, then “PUSHCONT” will be displayed (“PUSHCONT” will NOT be displayed if only calibrated with pH 4 or 10).
12. Push and release to continue or let the unit time out to exit after a 1-point or 2-point calibration.
13. Repeat steps 9 through 12 with pH 4 and 10 Buffer Solutions. After the 3rd calibration point is successfully saved, the display will indicate “SAVEd” and power off.
14. Verify calibration by retesting the calibration solution in solution check mode “SOL ck”, see section V below.

C. ORP Calibration using 80mV Quinhydrone, 260mV Quinhydrone, or 470mV MLC Light’s ORP Standard Solution.
NOTE: The PT4 has automatic temperature compensation in ORP calibration mode (from 15ºC to 30ºC).
1. Follow pH calibration steps 1 through 6, using ORP Solutions.
2. Release the button when “ORP” is displayed.
3. The display will indicate “CAL” and the LED will flash rapidly.
4. While the LED flashes rapidly, dip the PT4 in ORP Standard Solution so that the
sensor is completely submerged.
5. While the LED flashes slowly, the ORP calibration point will display along with “CAL”.
Swirl the PT4 around to remove any air bubbles, keeping the sensor submerged.
6. If the ORP calibration is successful, the display will indicate “CAL SAVEd”, then time out.
7. Verify calibration by retesting the calibration solution in solution check mode.

V. SOLUTION CHECK
Solution check is provided to verify the proper calibration value was recorded when using pH Buffers and ORP Standard Solutions. To verify proper calibration, simply put the PT4 into solution check mode, select the mode to verify (pH or ORP), then dip the sensor into the pH Buffer or ORP Calibration Solution so that the sensor is completely submerged and swirl around to release any air bubbles, then verify displayed value matches the value on the bottle.

To perform Solution Check:
1. Push and release the push button to turn the PT4 on.
2. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL”, “PAr SEL”, “SOL ck”, and “ESC”.
3. Release the button when “SOL ck” displays.
4. The display will alternate between “PUSHnHLD” and “SOL ck”.
5. Push and hold the button, The display will alternate between “PH” and “ORP”.
6. Release the button when desired mode (pH or ORP) is displayed.
7. While the LED flashes rapidly, dip the PT4 in FRESH buffer/calibration solution so that the sensor is completely submerged and swirl the PT4 around to remove any air bubbles.
8. Verify value displayed is correct.
NOTE: To verify ORP calibration while in solution check mode, you must manually correct for temperature variations from 25ºC. See instructions that come with the ORP Standard Solutions for temperature chart.

VI. Factory Calibration
When pH Buffers are not available, the PT4 can be returned to factory default calibration using the FAC CAL function. This will erase any stored wet calibration.
NOTE: Default factory calibration resets the electronics only and does NOT take the condition of the sensor into consideration.
To return your PT4 to factory calibration:
1. Push and release the push button.
2. Push and hold the button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL”, “PAr SEL”, “SOL ck”, and “ESC”.
3. Release the button when “FAC CAL” displays. The display will alternate between “PUSHnHLD” and “FAC CAL”.
4. Push and hold the push button. “SAVEd FAC” displays indicating the pen has been reset to its factory calibration.

MyronLMeters.com is the premier internet retailer of the Ultrapen PT4 and other reliable Myron L meters. Save 10% on Myron L meters when you order online HERE.

Categories : Application Advice, Care and Maintenance, Product Updates, Technical Tips

Ultrapen PT1 Conductivity, TDS and Salinity Calibration: Myron L Meters

Posted by 3 Apr, 2014

The Ultrapen PT1 is designed to be very reliable and requires only infrequent calibration. The factory recommends calibrating each measurement mode you use once monthly. However, you should check the calibration whenever measurements are not as expected. The PT1 is programmed for 2 calibration options: Wet Calibration or Factory Calibration. Wet calibration is most accurate. But if a high quality standard KCl-1800 µS or 442-3000 ppm solution is not available, the PT1 can be returned to factory settings.

A. Wet Calibration
Use calibration solution specified for measurement mode: Use KCL- 1800 for Cond KCl; Use 442-3000 for tdS 442, SALt 442, tdS NaCl, and SALt NaCl. See Specifications table for 442 solution ppm NaCl equivalent value. Calibrating TDS simultaneously calibrates SALT for the same value and vice versa.
1. Pour calibration solution into a clean container.
2. Rinse the pen 3 times by submerging the cell in fresh calibration solution and swirling it around.
3. Remove pen from solution, then fill the container one more time.
4. Press and release the push button. The LCD will briefly display the firmware version then the current measurement mode. Ensure the PT1 is in the correct solution mode.
5. Immediately push and hold the push button. The display will scroll through “CAL”, “SOL SEL”, “FAC CAL”, “ºCºF TEMP”, and “ESC”. Release the button when “CAL” displays.
6. Grasp the pen by its case with your fingers positioned between the
display and the pen cap to avoid sample contamination.
7. While the LED flashes rapidly, dip the pen in calibration solution so that the cell is completely submerged. If you do not submerge the cell in solution before the fl slows, allow the pen to power off and start over.
8. While the LED flashes slowly, swirl the pen around to remove bubbles, keeping the cell submerged. Keep pen at least 1 inch (2½ cm) away from sides/bottom of container.
9. When the LED light stays on solid, remove the pen from the solution. “CAL SAVED” will display indicating a successful calibration.
Note: If an incorrect solution is used or the measurement is NOT within calibration limits for any other reason, “Error” displays alternately with “CLEAn CEL/CHEC SOL”. Check to make sure you are using the correct calibration solution. If the solution is correct, clean the cell by submerging the cell in a 1:1 solution of Lime-A-Way® and water for 5 minutes. Rinse the cell and start over.
10. Small bubbles trapped in the cell can give a false calibration. Measure the calibration solution again to verify correct calibration. If the reading is not within ±1% of the calibration solution value, repeat calibration.

B. Factory Calibration
If you do not have the proper calibration solution or wish to restore the pen to its original factory settings for any other reason, use the FAC CAL function to calibrate the PT1.
1. Press and release the push button. The LCD will briefly display the firmware version then the current measurement mode.
2. Immediately push and hold the push button. The display will scroll through “CAL”, “SOL SEL”, “FAC CAL”, “ºCºF TEMP”, and “ESC”. Release the button when “FAC CAL” displays.
3. While the display scrolls through “PUSHnHLD” and “FAC CAL”, push and hold the push button until the display scrolls through “SAVEd” and “FAC CAL”, indicating the pen has been reset to its factory calibration.
4. Allow the pen to time out to turn power off.

MyronLMeters.com is the premier internet retailer of the Ultrapen PT1 and other reliable Myron L meters. Save 10% on Myron L meters when you order online HERE.

Categories : Product Updates

pH Calibration of the Ultrameter 6PFCE: MyronLMeters.com

Posted by 23 Mar, 2014

 

*Note: This procedure applies to the Ultrameter, PoolPro, TechPro, and D-6 Dialysate meter.

IMPORTANT: Always “zero” your Ultrameter II with a pH 7 buffer solution

before adjusting the gain with acid or base buffers, i.e., 4 and/or 10, etc.

a. pH Zero Calibration (6PFCE)

1. Rinse sensor well and cell cup 3 times with 7 buffer solution.

2. Refill both sensor well and cell cup with 7 buffer solution.

3. Press

pH

 

 

 

 

to verify the pH calibration. If the display shows 7.00, skip the pH

Zero Calibration and proceed to pH Gain Calibration.

4. Press

CAL key

 

 

 

 

 

to enter calibration mode. “CAL”, “BUFFER” and “7” will appear on the display.

display

 

 

 

 

 

 

 

Displayed value will be the uncalibrated sensor.

NOTES: If a wrong buffer is added (outside of 6-8 pH),“7” and “BUFFER

will flash, and the Ultrameter II will not adjust.

The uncalibrated pH value displayed in step 4 will assist in determining

the accuracy of the pH sensor. If the pH reading is above 8 with pH 7

buffer solution, the sensor well needs additional rinsing or the pH sensor

is defective and needs to be replaced.

5. Press

Up

 

 

 

 

or

Down

 

 

 

 

until the display reads 7.00.

NOTE: Attempted calibration of >1 pH point from factory calibration will

cause “FAC” to appear. This indicates the need for sensor replacement

or fresh buffer solution. The “FAC” internal electronic calibration is not intended to

replace calibration with pH buffers. It assumes an ideal pH sensor. Each “FAC”

indicates a factory setting for that calibration step (i.e., 7, acid, base).

You may press

CAL key

 

 

 

 

 

to accept the preset factory value, or you may

reduce your variation from factory setting by pressing

Up

 

 

 

 

or

Down

 

 

 

 

6. Press to accept the new value. The pH Zero Calibration

is now complete. You may continue with pH Gain Calibration or

exit by pressing any measurement key.

b. pH Gain Calibration (6PFCE)

IMPORTANT: Always calibrate or verify your Ultrameter II with a pH 7

buffer solution before adjusting the gain with acid or base buffers, i.e.,

4 and/or 10, etc. Either acid or base solution can be used for the 2nd

point “Gain” calibration and then the opposite for the 3rd point. The

display will verify that a buffer is in the sensor well by displaying either

Acd” or “bAS”.

1. The pH calibration mode is initiated by either completion of the

pH Zero Calibration, or verifying 7 buffer and pressing the

CAL key

 

 

 

 

 

key twice while in pH measurement mode.

2. At this point the “CAL”, “BUFFER” and “Acd” or “bAS

will be displayed (see Figures 7 and 8).

Capture

 

NOTE: If the “Acd” and “bAS” indicators are blinking, it indicates

an error and needs either an acid or base solution present in the sensor

well.

3. Rinse sensor well 3 times with acid or base buffer solution.

4. Refill sensor well again with same buffer solution.

5. Press

Up

 

 

 

 

or

Down

 

 

 

 

until the display agrees with the buffer value.

6. Press

CAL key

 

 

 

 

 

to accept the 2nd point of calibration. Now the

display indicates the next type of buffer to be used.

Single point Gain Calibration is complete. You may continue for the 3rd

point of Calibration (2nd Gain) or exit by pressing any measurement key.

Exiting causes the value accepted for the buffer to be used for both acid

and base measurements.

To continue with 3rd point calibration, use basic buffer if acidic buffer

was used in the 2nd point, or vice-versa. Again, match the display to the

known buffer value as in step 2 and continue with the following steps:

7. Repeat steps 3 through 6 using opposite buffer solution.

8. Press

CAL key

 

 

 

 

 

to accept 3rd point of calibration, which completes the Calibration procedure.

Fill sensor well with Sensor Storage Solution and replace protective cap.

You can find technical advice and videos, the calibration solutions you need, and reliable Myron L meters
at MyronLMeters.com
Categories : Application Advice, Care and Maintenance, Product Updates, Technical Tips

TDS Calibration on the Ultrameter II 6PIIFCe: MyronLMeters.com

Posted by 23 Mar, 2014

 

*Note: This procedure applies to the Ultrameter, PoolPro, TechPro, and D-6 Dialysate meter.

a. Fill and rinse the conductivity cell three times with a 442 standard solution. In this example, we’re using 442-3000.

b. Refill conductivity cell with same standard solution you rinsed with.

c. Press

TDS

 

 

 

 

 

then press

 

.CAL key

 

 

 

 

The “CAL” icon will appear in the top center of the display. In this example, the reading shows 2988.

d. Press

Up

 

 

 

 

 

or

Down

 

 

 

 

 

to step the displayed value toward the standard’s value.

In this example, we’re pressing

 

Up

 

 

 

 

 

to go down from 2988 to 3000. You can also hold a key down to scroll rapidly.

e. Press

 

CAL key

 

 

 

 

 

once to confirm the new value and end the calibration.

You can find technical advice and videos, the calibration solutions you need, and reliable Myron L meters
at MyronLMeters.com
 
 

 

Categories : Application Advice, Care and Maintenance, Product Updates, Technical Tips

Conductivity Calibration on the Ultrameter II 6PIIFCe: MyronLMeters.com

Posted by 23 Mar, 2014

 

*Note: This procedure applies to the Ultrameter, PoolPro, TechPro, and D-6 Dialysate meter.

a.  Fill and rinse the conductivity cell three times with a KCL standard solution. In this example, we’re using KCL-7000.

b. Refill conductivity cell with same standard solution you rinsed with.

c. Press

COND

 

 

 

 

then press

CAL key

.

 

 

 

The “CAL” icon will appear on the display.

display

 

 

 

 

 

 

 

 

 

 

 

d. Press

Up

 

 

 

or

Down

 

 

 

 

to step the displayed value toward the standard’s value.

In this example, we’re pressing

Down

 

 

 

 

to go down from 7032 to 7000. You can also hold a key down to scroll rapidly.

e. Press

CAL key

 

 

 

 

 

once to confirm the new value and end the calibration.

You can find technical advice and videos, the calibration solutions you need, and reliable Myron L meters
at MyronLMeters.com
Categories : Application Advice, Care and Maintenance, Product Updates, Technical Tips

Ultrapen PT2 Calibration – MyronLMeters.com

Posted by 28 Oct, 2012


Ultrapen PT2 Calibration
We recommend calibrating twice a month, depending on usage. However, you should check the calibration whenever measurements are not as expected. 3-point Wet Calibration is most accurate and is recommended.

NOTE: If the measurement is NOT within calibration limits for any reason, “Error” will display. Check to make sure you are using a proper pH buffer solution. If the solution is correct, clean the glass bulb of the sensor with a cotton swab soaked in isopropyl alcohol. Restart calibration.

NOTE: Small bubbles trapped in the sensor may give a false calibration. After calibration is completed, measure the pH buffer solutions again to verify correct calibration.

NOTE: If at any point during calibration, you do not submerge the sensor in solution
before the flashing slows, allow the pen to power off and start over.

A. Calibration Preparation
1. For maximum accuracy, fill 2 clean containers with each pH buffer. Arrange them in such a way that you can clearly remember which is the rinse solution and which is the calibration buffer. If you don’t have enough buffer, you can use
1 container of each buffer for calibration and 1 container of clean water for all rinsing. Always rinse the pH sensor between buffer solutions.
2. Ensure the pH sensor is clean and free of debris.

B. 3-Point Calibration
Use pH 7, 4 and 10 buffers for 3-point calibration. You can find the buffer solutions here: http://www.myronlmeters.com/pH-Buffer-Calibration-Solutions-s/82.htm
1. Thoroughly rinse the pen by submerging the sensor in pH 7 buffer rinse solution and swirling it around.
2. Push and release the push button to turn the unit on.
3. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
4. Release the button when “CAL” displays. The display will indicate “CAL” and
the LED will flash rapidly.
5. While the LED flashes rapidly, dip the pen in pH 7 buffer calibration solution
so that the sensor is completely submerged.
6. While the LED flashes slowly, the pH calibration point will display along with
“CAL”. Swirl the pen around to remove bubbles, keeping the sensor submerged.
7. If the pH 7 calibration is successful, the display will indicate “SAVEd”, then
“PUSHCONT” will be displayed.
8. Push and release the push button to continue. The LED will begin flashing rapidly.
9. Repeat steps 5 through 8 with pH 4 and 10 buffer calibration solutions.
10. After the 3rd calibration point is successfully saved, the display will indicate
“SAVEd” and power off.
11. Verify calibration by retesting the calibration solution.

C. 2-Point Calibration
Use pH 7 and 4 or 10 buffers for 2-point calibration.
1. Thoroughly rinse the pen by submerging the sensor in pH 7 buffer rinse solution and swirling it around.
2. Push and release the push button to turn the unit on.
3. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
4. Release the button when “CAL” displays. The display will indicate “CAL” and
the LED will flash rapidly.
5. While the LED flashes rapidly, dip the pen in pH 7 buffer calibration solution
so that the sensor is completely submerged.
6. While the LED flashes slowly, the pH calibration point will display along with
“CAL”. Swirl the pen around to remove bubbles, keeping the sensor submerged.
7. If the pH 7 calibration is successful, the display will indicate “SAVEd”, then
“PUSHCONT” will be displayed.
8. Push and release the push button to continue. The LED will begin flashing rapidly.
9. Repeat steps 5 through 7 with pH 4 or 10 buffer calibration solution.
10. Leave the pen in the same buffer solution until the unit powers off. The offset will be applied to the remaining calibration point.
11. Verify calibration by retesting the calibration solution.

D. 1-Point Calibration
Use pH 7, 4 or 10 buffer for 1-point calibration.
1. Thoroughly rinse the pen by submerging the sensor in pH buffer rinse solution and swirling it around.
2. Push and release the push button to turn the unit on.
3. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
4. Release the button when “CAL” displays. The display will indicate “CAL”
and the LED will flash rapidly.
5. While the LED flashes rapidly, dip the pen in pH buffer calibration solution
so that the sensor is completely submerged.
6. While the LED flashes slowly, the pH calibration point will display along with “CAL”;
swirl the pen around to remove bubbles, keeping the sensor submerged.
7. If the pH calibration is successful, the display will indicate “SAVEd”, then “PUSHCONT” will be displayed. “PUSHCONT” will not display if you calibrated 4 or 10.
8. Leave the pen in the same buffer solution until the unit powers off. The offset will be applied to the remaining calibration points.
9. Verify calibration by retesting the calibration solution.

E. Factory Calibration
When pH buffers are not available, the PT2 can be returned to factory default calibration using the FAC CAL function. This will erase any stored wet calibration. NOTE: default factory calibration resets the electronics only and does NOT take the condition of the sensor into consideration.
To return your unit to factory calibration:
1. Push and release the push button.
2. Push and hold the push button. The display will alternate between “CAL”, “FAC CAL”, “ºCºF TEMP”, “ModE SEL” and “ESC”.
3. Release the button when “FAC CAL” displays. The display will alternate between “PUSHnHLD” and “FAC CAL”.
4. Push and hold the push button. “SAVEd FAC” displays indicating the pen has been reset to its factory calibration.

Myron L Meters is the premier online retailer of accurate, reliable, easy-to-use Myron L meters.

Please visit us on the web at:

http://www.myronlmeters.com

Facebook:

http://www.facebook.com/pages/Myron-L-Meters/147455608645777

Twitter:

http://twitter.com/MyronLMeters

Google +:

https://plus.google.com/i/0gw_uw5P328:dbK-UM_4xek

Linkedin:

http://www.linkedin.com/profile/view?id=98473409&trk=tab_pro

YouTube:

http://www.youtube.com/myronlmeters

News:

http://waterindustrynews.com

Categories : Product Updates, Technical Tips

Myron L Meters Thanks Alliance Calibration!

Posted by 18 Jul, 2011
Scientist

Alliance Calibration is centrally located in Cincinnati, Ohio.

Our client base includes Aeronautical, Defense, Automotive, Government, Research, Medical, Pharmaceutical, Energy & Power industries.

Do you have a need for rapid turnaround on your calibrations? 
Rapid turnaround is not a problem at Alliance Calibration. We ship daily to the entire United States.

Are on-site calibrations critical to your operation?
We have the capabilities and expertise to perform most calibrations on-site at your facility to prevent costly down time. We are accredited to ISO:17025 and our calibration staff have earned American Society for Quality certifications.

Do you need access to your calibration certificates 24 hours per day, 365 days per year
Our eTracking program allows you access your data from your computer 24 hours a day! In addition, you can print your calibration certificates on your own printer!

More about this fine company here: http://www.alliancecalibration.com/

There’s always more at MyronLMeters.com.

Categories : Uncategorized

Myron L Meters Thanks Alliance Calibration!

Posted by 18 Jul, 2011
Scientist

Alliance Calibration is centrally located in Cincinnati, Ohio.

Our client base includes Aeronautical, Defense, Automotive, Government, Research, Medical, Pharmaceutical, Energy & Power industries.

Do you have a need for rapid turnaround on your calibrations? 
Rapid turnaround is not a problem at Alliance Calibration. We ship daily to the entire United States.

Are on-site calibrations critical to your operation?
We have the capabilities and expertise to perform most calibrations on-site at your facility to prevent costly down time. We are accredited to ISO:17025 and our calibration staff have earned American Society for Quality certifications.

Do you need access to your calibration certificates 24 hours per day, 365 days per year
Our eTracking program allows you access your data from your computer 24 hours a day! In addition, you can print your calibration certificates on your own printer!

More about this fine company here: http://www.alliancecalibration.com/

There’s always more at MyronLMeters.com.

Categories : Uncategorized

Myron L Products

Posted by 19 Dec, 2014

Part Number   Description                                 Price
0241-1 024-1: 1″ NPT 24 VAC Solenoid Valve 213
35BLBO 35BLBO 3.5 Digit Backlit Liquid Crystal Display (LCD) with Bezel 228
35BLO 35BLO 3.5 Digit Backlit Liquid Crystal Display (LCD) 228
3RE 3RE 3 Range Switch with Enclosure (Add C=Cond or R=Res) 248.5
3RH 3RH Harness, 3RO/3SRO 3 ft./1 meter Multi-Wire Ribbon Cable 44.5
3RMO 3RMO Module only, 3RO 3 Range 97
3RO 3RO 3 Range Switch with 1 meter/3 ft. cable (Add C=Cond or R=Res)* 133
3SH 3SH Harness, 3SO 3 ft./1 meter 5-Wire Sensor Cable 32
3SMO 3SMO Module only, 3SO 3 SENSOR Input 107
3SO 3SO: 3 Sensor Input Module with 3 ft./1 meter cable (customer installed) 133
3SRE 3SRE Above 3S and 3R combined with Enclosure (Add C=Cond or R=Res) 280.5
3SRH 3SRH Harness, 3SO 3 ft./1 meter 2-wire Disable 29
3SRMO 3SRMO Module only, 3SRO 3 SENSOR Input/3 Range 129
3SRO 3SRO Above 3S and 3R combined (Add C=Cond or R=Res)* 181.5
45BLBO 45BLBO 4.5 Digit Backlit Liquid Crystal Display (LCD) with Bezel 256
45BLO 45BLO 4.5 Digit Backlit Liquid Crystal Display (LCD) 256
4AH 4AH Harness, 4A (4-20 mA) 47
4AMO 4AMO Module only, 4-20 mA Self-Powered Isolated Output (OEM & digital Monitor/controllers only) 154.5
4AO 4AO: 4-20 mA Self-Powered Isolated Output Module with harness (digital models only) 170.5
597 CONTROLSTIK II SYSTEM 597 837
9P 9P Instrument only; Conductivity, Resistivity, TDS, Alkalinity, Hardness, Saturation Index,ORP/Free Chlorine, pH, Temperature 1349
35CCPP Solutions Sampling Syringe: 35cc 7.5
4AH Harness, 4A (4-20mA) 170.5
4AMO 4-20 mA Self-Powered Isolated Output Module 154.5
ATT100 Alkalinity Test Tablets: ATT100 120.5
BD BluDock Wireless Data Transfer Module: BD 208
BD-0001 BluDock Wireless Data Transfer Module: BD – No options available for this product 208
BDDO Bluetooth USB adapter: BDDO 25
CLK Litho-Kit: CLK 166.5
D-1P Replacement Plunger D-1 17.5
DOM5 DO Membrane 5pk forUltrapen PT5 TBD
DOM5K DO Membrane 5pk, Replenishment Electrolyte 2 oz, Syringe w/Tip Kit TBD
DORS DO Replenishment Syringe w/Tip for PT5 TBD
DOSRE DO Sensor Replenishment Electrolyte 2 oz for PT5 TBD
NIST NIST Certificate (may delay shipment) 42
NIST-Field NIST Certificate Field 83
PK3 Porta-Kit: PK3 103
PK7 Porta-Kit: PK7 103
PKD Porta-Kit: PKD 146.5
PKP Porta-Kit: PKP 126
PKPS Porta-Kit: PKPS 152
PKUU Porta-Kit: PKUU 143.5
PLK Litho-Kit: PLK 177
PSTK Soil Test Kit: PSTK 143.5
PTP Porta-Pack: PTP 58
PTSC Replacement sensor cap – Ultrapen 5.5
RE-10 Range Extender: RE-10 20
RPG Replacement pH Sensor: RPG 154
RPHC6 Replacement pH cap, Ultrameter 13.5
RPR Replacement pH & ORP Sensor: RPR 179
RPT2 Replacement pH Sensor PT2: RPT2 50
RPT3 Replacement ORP Sensor PT3: RPT3 89
RPT4 Replacement Free Chlorine Sensor PT4: RPT4 119
RPY Replacement pH Sensor: RPY 144.5
SRW6321 Replacement screws and O Ring 2
STK Soil Test Kit: STK 115.5
TKPS9A TKPS9A Replacement AHL Titration Kit for PS9 includes hard protective case with all required accessories;no instrument 360.5
UCC Canvas Case: UCC 42
UCCDT Canvas Case: UCCDT 42
ULK Litho-Kit: ULK 215.5
UPP Porta-Pack: UPP 66.5
VBP Valve Bleed Plug for 024-1 Solenoid Valve 9
512M10 512M10: 0-10,000 micromhos/microsiemens 209.5
512M2 512M2: 0-50 micromhos/microsiemens 209.5
512M3 512M3: 0-500 micromhos/microsiemens 233
512M4 512M4: 0-2500 micromhos/microsiemens 209.5
512M5 512M5: 0-5000 micromhos/microsiemens 209.5
512T10 512T10: 0-10,000 ppm 209.5
512T2 512T2: 0-50 ppm 209.5
512T3 512T3: 0-500 ppm 209.5
512T4 512T4: 0-2500 ppm 209.5
512T5 512T5: 0-5000 ppm 209.5
512T5D 512T5D: 0-5000 ppm (442 & NaCL) 218.5
532M1 532M1: 0-50, 500, 5000 micromhos/microsiemens 328
532T1 532T1: 0-50, 500, 5000 ppm 328
532T2 532T2: 0-25, 250, 2500 ppm 328
AG-5 Agri-Meter – Ag-5: 0-5 millimhos 209.5
AG6/PH Agri-Meter – Ag-6: 0-5 millimhos; 2-12 pH 591.5
D-1 Dialysate Meter – D-1: 11-15 millimhos 263.5
D-2 Dialysate Meter – D-2: 0-10 & 0-20 millimhos 383
EP EP: 0-0.5, 5, 50, 500, 5000 micromhos/microsiemens, 2-30 megohms 369.5
EP-10 EP-10: 0-10, 100, 1000, 10,000 micromhos/microsiemens 357.5
EP11/PH EP-11/PH: 0-10, 100, 1000, 10,000 micromhos/microsiemens; 2-12 pH 566.5
M6/PH M6/PH: 0-5000 micromhos/microsiemens; 2-12 pH 467
RO-1 RO Meter – RO-1: 0-1250 ppm with color band 209.5
RO-1NC RO Meter – RO-1: 0-1250 ppm without color band 209.5
T2/PH T2/PH: 0-50, 500, 5000 ppm; 2-12 pH 556
T6/PH T6/PH: 0-5000 ppm; 2-12 pH 461.5
ALK-100 ALK-100 Alkalinity Standard Solution; sodium bicarbonate as ppm calcium carbonate;available in 2oz./Qt 18
ATK ATK Alkalinity Test Kit: 1 vial and 20 tablets 44.5
ATT ATT Alkalinity Test Tablets, package of 1000 1206.5
ATT100 ATT100 Alkalinity Test Tablets, package of 100 120.5
BATT11768 BATT11768 Replacement N type battery 2
BATT11768-B3 BATT11768-B3 Replacement N type battery in QTY: 3 5.5
BDDO BDDO Bluetooth USB adapter (dongle) 25
BLUDOCK BLUDOCK bluDock Wireless Data Transfer Module, U2CI software CD, wireless USB adapter (dongle) and instructions for models 4PII (S/Ns 4220001 and up) 208
C4AO C4AO 4-20 mA Output Module (when ordered after initial purchase) 144.5
CC or RC CC or RC NIST Certificate without sensor (C=Cond or R=Res) 53
CCATF or RCATF CCATF or RCATF NIST Certificate without sensor, ordered after instrument has shipped (C=Cond or R=Res) 83
CCS CCS NIST Certificate with sensor, using Standard Solution (Conductivity/TDS only) 119.5
CLTMO CLTMO Limit Timer Module with small Knob (when ordered after initial purchase) 106
D-4 Digital Dialysate Meter – D-4 1306.5
D-6 Digital Dialysate Meter – D-6 1587.5
PS9 Pool Pro PS9 (Kit not included – Instrument only) 1214.1
PS9-TKA Pool Pro PS9TKA 1407.15
TH1 Techpro II – TH1: DISCONTINUED 392.5
TP1 Techpro II – TP1 607.5
TPH1 Techpro II – TPH1 841
4PII Ultrameter II – 4PII 709
6PII Ultrameter II – 6PII 947.5
6PIIFC Ultrameter II – 6PIIFC 947.5
6PIIFCe Ultrameter II – 6PIIFCe 947.5
PS6 POOLPRO – PS6 947.5
9PTK Ultrameter III – 9PTK 1512.5
9PTKA Ultrameter III – 9PTKA 1512.5
PT1 UltraPen – PT1 Conductivity, TDS, Salinity Pen 147
PT1-0001 UltraPen – PT1 Conductivity, TDS, Salinity Pen – No options available for this product 147
PT1-B24 UltraPen – PT1 Conductivity, TDS, Salinity Pen (24 pack) 3391.2
PT2-SS UltraPen – PT2 pH and Temperature Pen with pH Storage Solution 139.5
PT2-SS-B24 UltraPen – PT2 pH and Temperature Pen with pH Storage Solution (24 Pack) 3110.4
PT3 UltraPen – PT3 ORP/REDOX and Temperature Pen 167.5
PT3-B24 UltraPen – PT3 ORP/REDOX and Temperature Pen (24 pack) 3866.4
PT4 UltraPen – PT4 Free Chlorine and Temperature Pen 242.1
PT5 UltraPen – PT5 Dissolved Oxygen and Temperature Pen 0
PTBT1 UltraPen – PTBT1 Conductivity, TDS, Salinity Pen – Bluetooth 311.4
PTBT2 UltraPen – PTBT2 pH and Temperature Pen Bluetooth 333
PTCOMBO UltraCombo – PT1 & PT2 Conductivity, TDS, Salinity, pH, Temp Pens 280.5
PTCOMPLETE ULTRAPEN Complete – PT1, PT2, PT3, PT4 – Conductivity, TDS, Salinity, pH, ORP, Free Chlorine, Temp Pens 679.5
PTSET ULTRAPEN Set – PT1, PT2, & PT3 – Conductivity, TDS, Salinity, pH, ORP, Temp Pens 448.5
PTSPLASH ULTRAPEN Splash – PT1, PT2, PT4 – Conductivity, TDS, Salinity, pH, Free Chlorine, Temp Pens 528.75
EDTA-HC EDTA-HC Hardness reagent H3; high concentration ethylenediaminetetraacetic acid ?10%wv;available in 2oz./Qt 18
EDTA-LC EDTA-LC Hardness reagent H2; low concentration ethylenediaminetetraacetic acid ?5%wv;available in 2oz./Qt 18
F100 F100 Fuse, 115/230 VAC control board, 100 mA (T.10A) 5 X 20 slow blow/time delay 5
F250 F250 Fuse, 24 VAC Main Control Board, 250 mA (T.25A) 5 X 20 Slow Blow/Time Delay 5
FS4 FS4 Flow Switch 153.5
FSM FSM Flow Switch and Manifold 305
FSMP FSMP 15 x 20 in./381 x 508 mm Mounting Panel (horizontal or vertical mounting) 83
FVMP-100 FVMP-100 M’PET fixed volume 100microL pipette 37.5
GTCHB412M2415PC GTCHB Series Band Clamp Filter Housings – 4 Cartridge – 304SS – 10inch 754
H2SO4-1 H2SO4-1 Alkalinity reagent A1; sulfuric acid <1%wv; available in 2oz./Qt 18
HARD-200 HARD-200 Hardness Standard Solution; calcium chloride as ppm calcium carbonate;available in 2oz./Qt 18
HCL-3 HCL-3 Hardness reagent C1; hydrochloric acid <3%wv; available in 2oz./Qt 18
HI96106 HI 96106 – (Champ) Economical pH Tester with 0.2 Accuracy 35
241 024-1: 1″ NPT 24 VAC Solenoid Valve 213
3RE 3RE: Three Range Switch Module 248.5
3SE 3SE: Three Sensor Input Module 248.5
3SRE 3SRE: Three Sensor Input/ Three Range Module 280.5
721II 721II: pH Analog Monitor Only 523
726II 726II: ORP Analog Monitor Only 523
751II 751II: Resistivity Analog Monitor Only 523
756II 756II: Conductivity Analog Monitor Only 508
722II 722II: pH Analog Monitor/Controller 600
727II 727II: ORP Analog Monitor/Controller 600
752II 752II: Resistivity Analog Monitor/Controller 600
757II 757II: Conductivity Analog Monitor/Controller 585.5
ASI Aquaswitch I 583.5
CS CS: RESISTIVITY CALIBRATION MODULE 56
CS51 CS51: Conductivity Sensor 228.5
CS51LC CS51LC: Low-Range Conductivity Sensor 228.5
CS51LS CS51LS: Long Conductivity Sensor 313
CS51R CS51R: Conductivity Sensor 247.5
CS52 CS52: High-Range Conductivity Sensor 267
CS11 CS-11: 20 Megohms Resistivity Calibration Module 56
CS12 CS-12: 10 Megohms Resistivity Calibration Module 56
CS13 CS-13: 5 Megohms Resistivity Calibration Module 56
CS14 CS-14: 2 Megohms Resistivity Calibration Module 56
CS15 CS-15: 1 Megohms Resistivity Calibration Module 56
CS16 CS-16: 500 Kilohms Resistivity Calibration Module 56
CS17 CS-17: 200 Kilohms Resistivity Calibration Module 56
724II 724II: pH Digital Monitor Only 494.5
729II 729II: ORP Digital Monitor Only 494.5
754II 754II: Resistivity Digital Monitor Only 494.5
759II 759II: Conductivity Digital Monitor Only 479
723II 723II: pH Digital Monitor/Controller 570.5
728II 728II: ORP Digital Monitor/Controller 570.5
753II 753II: Resistivity Digital Monitor/Controller 570.5
758II 758II: Conductivity / TDS Digital Monitor/Controller 555
O72D O72D: 1/2″ MNPT Double Junction 297.5
O72F O72F: 1/2″ MNPT Heavy Duty FLAT TIP Double Junction 370
O72LC O72LC: 1/2″ MNPT Double Junction (Low Conductivity) 316
O72S O72S: 1/2″ MNPT Single Junction 285
O74DR O74DR: 3/4″ MNPT Double Junction 297.5
O74FR O74FR: 3/4″ MNPT Heavy Duty FLAT TIP Double Junction 370
O74LCR O74LCR: 3/4″ MNPT Double Junction (Low Conductivity) 316
O74SR O74SR: 3/4″ MNPT Single Junction 285
P72D P72D: 1/2″ MNPT Double Junction 282.5
P72F P72F: 1/2″ MNPT Heavy Duty FLAT TIP Double Junction 356.5
P72LC P72LC: 1/2″ MNPT Double Junction (Low Conductivity) 300.5
P72S P72S: 1/2″ MNPT Single Junction 268.5
P74DR P74DR: 3/4″ MNPT Double Junction 282.5
P74FR P74FR: 3/4″ MNPT Heavy Duty FLAT TIP Double Junction 356.5
P74LCR P74LCR: 3/4″ MNPT Double Junction (Low Conductivity) 300.5
P74SR P74SR: 3/4″ MNPT Single Junction 268.5
RA Remote Alarm 235.5
CS10 CS10: Resistivity Sensor 296.5
CS10LS CS10LS: Long Resistivity Sensor 378.5
SCO SCO: Second Alarm/control Module with harness (Monitor/controllers only) 151.5
SMP50 SMP50: Surface Mounting Plate 29
VR VR: Voltage Regulator (115 VAC down to 24 VAC) 50
JM4P JM4P Jumper, in place of SC Module 20
MBV MBV Mini Ball Valve (Sample Port) 27
MCRD-(length) MCRD-(length) Cable, 5-Wire Sensor Cable (~ $3.23/meter) 1.5
NAOH NAOH Hardness reagent H1; sodium hydroxide <4%wv; available in 2oz./Qt 18
OC OC: NIST Certificate ORP without sensor 53
ORPCOND “ORPCOND ORP cell conditioner solution; sulfuric acid, ferrous ammonium sulfate,
ferric ammonium sulfate” 18
PAO PAO ~70dB Piezo Electric Alarm only (OEM and digital Monitor/controllers only) 51
PATO PATO ~70dB Piezo Alarm and Timer Module (OEM and digital Monitor/controllers only) 179.5
PCO PCO: Power Cord 115 VAC (8 ft./2,44 meters with plug and strain relief) – Not for 230 VAC use 26
PHC PHC: NIST Certificate pH without sensor 53
PHCRD PHCRD Cable, 5-Wire pH Sensor Cable (~$3.25/meter) 3.5
PHCRD-(length) PHCRD-(length) Cable, 5-wire pH Sensor Cable (~$3.25/meter) 3.5
PHCS PHCS: NIST Certificate pH with sensor, using Buffer Solutions 119.5
PROKIT PROKIT Quantity of 4 micropipettes with 48 replacement tips 154
PS9 “PS9 Instrument only; Conductivity, Mineral/Salt Concentration, TDS, Alkalinity, Hardness, Saturation
Index, ORP/Free Chlorine, pH, Temperature ” 1349
PS9-BD PS9-BD Instrument only; Conductivity, Salt, TDS, Alkalinity, Hardness, SaturationIndex, ORP/Free Chlorine, pH, Temperature 1557
PS9TK PS9TK PS9TK AHL Titration Kit 1463.5
PTIP100-48 PTIP100-48 Replacement tips for M’PET micropipette (bag of 48) 5
PTSC RPT2 Replacement Sensor Cap with foam installed 5.5
RAH RAH Harness, female Methode connector with 8 in./203 mm 2 conductor wire.For use with RA, PA, 3S (Flow Switch), etc. 18.5
RM-16 RM-16: 0-500 kilohm a range module for 750II resistivity 24.5
RM3R RM3R Range Module, 3RMO/3SRMO Resistivity only 26
442-1000G 442-1000 (1000 ppm) 1 gallon 80
442-1000Q 442-1000 (1000 ppm) 32 oz 23.5
442-1500-2OZ 442-1500 (1500 ppm) 2 oz 10
442-15000G 442-15000 (15000 ppm) 1 gallon 80
442-15000Q 442-15000 (15000 ppm) 32 oz 23.5
442-1500G 442-1500 (1500 ppm) 1 gallon 80
442-1500Q 442-1500 (1500 ppm) 32 oz 23.5
442-150G 442-150 (150 ppm) 1 gallon 80
442-150Q 442-150 (150 ppm) 32 oz 23.5
442-15G 442-15 (15 ppm) 1 gallon 80
442-15Q 442-15 (15 ppm) 32 oz 23.5
442-3000-2OZ 442-3000 (3000 ppm) 2 oz 10
442-30000G 442-30000 (30000 ppm) 1 gallon 80
442-30000Q 442-30000 (30000 ppm) 1 quart 23.5
442-3000G 442-3000 (3000 ppm) 1 gallon 80
442-3000Q 442-3000 (3000 ppm) 1 quart 23.5
442-300G 442-300 (300 ppm) 1 gallon 80
442-300Q 442-300 (300 ppm) 32 oz 23.5
442-30G 442-30 (30 ppm) 1 gallon 80
442-30Q 442-30 (30 ppm) 32 oz 23.5
442-500Q 442-500 (500 ppm) 32 oz 23.5
DDD6-SOL-SET Digital Dialysate Meter D6 Calibration Solution Set (6 bottles 32oz) 141
EDTA-HC-2oz EDTA-HC Hardness reagent H3 Solutions: 2 oz 17.5
EDTA-HC-QT EDTA-HC Hardness reagent H3 Solutions: QT 33.5
EDTA-LC-2oz EDTA-LC Hardness reagent H2 Solutions: 2 oz 17.5
EDTA-LC-QT EDTA-LC Hardness reagent H2 Solutions: QT 33.5
H2SO4-2oz H2SO4-1 Alkalinity reagent A1 Solutions: 2 oz 17.5
H2SO4-QT H2SO4-1 Alkalinity reagent A1 Solutions: QT 33.5
HCL3-QT HCL-3 Hardness reagent C1 Solutions: QT 33.5
KCL-1800-2OZ KCL-1800 (1800 uS) 2 oz 10
KCL-18000G KCL-18000 (18000 uS) 1 gallon 80
KCL-18000Q KCL-18000 (18000 uS) 32 oz 23.5
KCL-1800G KCL-1800 (1800 uS) 1 gallon 80
KCL-1800Q KCL-1800 (1800 uS) 32 oz 23.5
KCL-180G KCL-180 (180 uS) 1 gallon 80
KCL-180Q KCL-180 (180 uS) 32 oz 23.5
KCL-18G KCL-18 (18 uS) 1 gallon 80
KCL-18Q KCL-18 (18 uS) 32 oz 23.5
KCL-7000-2OZ KCL-7000 (7000 uS) 2 oz 10
KCL-70000G KCL-70000 (70000 uS) 1 gallon 80
KCL-70000Q KCL-70000 (70000 uS) 32 oz 23.5
KCL-7000G KCL-7000 (7000 uS) 1 gallon 80
KCL-7000Q KCL-7000 (7000 uS) 32 oz 23.5
KCL-700G KCL-700 (700 uS) 1 gallon 80
KCL-700Q KCL-700 (700 uS) 32 oz 23.5
KCL-70G KCL-70 (70 uS) 1 gallon 80
KCL-70Q KCL-70 (70 uS) 32 oz 23.5
NACL-12.5G NACL-12.5 (12.5 mS) 1 gallon 80
NACL-12.5Q NACL-12.5 (12.5 mS) 32 oz 23.5
NACL-13.4G NACL-13.4 (13.4 mS) 1 gallon 80
NACL-13.4Q NACL-13.4 (13.4 mS) 32 oz 23.5
NACL-14.0G NACL-14.0 (14.0 mS) 1 gallon 80
NACL-14.0Q NACL-14.0 (14.0 mS) 32 oz 23.5
NACL-7500G NACL-7500 (7500 ppm) 1 gallon 80
NACL-7500Q NACL-7500 (7500 ppm) 32 oz 23.5
NAOH-2oz NAOH Hardness reagent H1 Solutions: 2 oz 18
NAOH-QT NAOH Hardness reagent H1 Solutions: QT 35
ORP2602oz ORP 260mV Quinhydrone kit Solutions: 2 oz 16.5
ORP4702oz ORP 470mV MLC Light’s ORP Calibration Solution: 2 oz 16.5
ORP802oz ORP 80mV Quinhydrone Kit Solutions: 2 oz 16.5
ORPCALKIT ORP Calibration Kit: solutions and powders for 80, 260, 470mV 41.5
ORPCALKIT-SSQ Ultrapen PT3 / PT4 ORP Calibration Kit (2oz) for 80, 260, 470mV + pH Sensor Storage Solution (32oz) 65
ORPCOND-QT ORP cell conditioner solutions: QT 94.5
PH10-2OZ pH 10 Buffer: 2 oz 10.5
PH10G pH 10 Buffer: 1 gallon 80
PH10Q pH 10 Buffer: 32 oz 23.5
PH4-2OZ pH 4 Buffer: 2 oz 10.5
PH4G pH 4 Buffer: 1 gallon 80
PH4Q pH 4 Buffer: 32 oz 23.5
PH7-2OZ pH 7 Buffer: 2 oz 10.5
PH7G pH 7 Buffer: 1 gallon 80
PH7Q pH 7 Buffer: 32 oz 23.5
PPPS6-SOL-SET PoolPro PS6 Calibration Solution Set (6 bottles 32oz) KCL-7000, 442-3000, pH Buffer 4, pH Buffer 7, pH Buffer 10, pH Sensor Storage Solution 141
SSG pH/ORP Sensor Storage Solutions: 1 gallon 83
SSQ pH/ORP Sensor Storage Solutions: 32 oz 23.5
UM4PII-SOL-SET Ultrameter II 4P Calibration Solution Set (2 bottles 32oz) – KCL-7000Q, 442-3000Q 48
UM6P-SOL-SET Ultrameter 6P Calibration Solution Set (6 bottles 32oz) KCL-7000, 442-3000, pH Buffer 4, pH Buffer 7, pH Buffer 10, pH Sensor Storage Solution 141
UPPT1-SOL-SET Ultrapen PT1 Calibration Solution Set (2 bottles 32oz) – KCL-7000Q, 442-3000Q 48
UPPT2-SOL-SET Ultrapen PT2/TH1 Calibration Solution Set (4 bottles 32oz) – pH Buffer 4, pH Buffer 7, pH Buffer 10, and pH Sensor Storage Solution 95
S24 24 VAC Solenoid Only for 024-1 Valve 40.5
SCH SCH Harness, Second Alarm/control 65
SCMO SCMO Second Alarm/control Module (Monitor/controllers only) 114.5
SCO SCO Second Alarm/control Module with harness (Monitor/controllers only) 151
SQCO SQCO Quik-Connect, sensor cable M/F with 12 in./305 mm lead 69
TDH TDH Harness, TP and TPA (Temperature Module) 64
TFM4 TFM4 T Fitting 3/4 SCH 80 (for use with CSAR Sensor and Mini Ball Valve) 42
THO THO TPM/TPMO Alarm/control Harness only (requires SCO and TPMO options) 77
TK9 “TK9 Replacement AHL Titration Kit for 9PTKA includes hard protective case with all required accessories;
no instrument ” 350.5
TKCE TKCE Cell cup extender with tethered cap 26
TPAMO TPAMO Above TP with Adjustable Hysteresis (Special Order) 167.5
TPC TPC Calibration Module (Temperature Module) 168.5
TPLUNGER TPLUNGER Titration plunger 18.5
TPMH TPMH 18 in./457 mm Harness, OEM Temperature Module 44.5
TPMO TPMO Module only with 18 in./457 mm OEM Harness, Requires Temperature Sensor (ITS1), Temperature 0-200’C (OEM and digital Monitor/controllers only) 165.5
TPO TPO: Temperature Module 0-200¡C with harness (digital models only, requires temperature sensor ITS1) 199
U2CIP U2CIP uDock� Accessory Package for original models 4PII, 6PII and PS6 208

Categories : Product Updates

The Ultrameter II 6P

Posted by 9 Dec, 2014

 

 

Myron L Ultrameter II 6P

Myron L Ultrameter II 6P

 

 

 

 

 

 

 

 

 

 

As featured in: Environmental-Expert.com, Greenhouse Product News, Environmental Technology Online, Water and Wastes Digest

“The Myron L Ultrameter is a terrific product – I have used them for years.
The product is robust and durable.”
Larry L. Managing Director

Quick, Accurate Readings April 28, 2014

Reviewer: Manjit M, Project Engineer from UAE

Me & all my team uses ULTRAMETER II almost daily. It’s a great piece of equipment for Quick Accurate Readings. This handheld device is very easy to use & calibrate. Many times I use the hand held device as standard reference for my online meters, Another attractive part is the durability of the equipment – it will last for quite good period of time. Myron L products are undoubtedly reliable

Great Product, One of the Best! April 27, 2014

Reviewer: Env. Tech, Academic/Professional Researcher from California

I use this handheld many times each day. We have at least 3 on site at all times for regulatory purposes. I get accurate readings all the time in many types of water. It calibrates easier than any other handheld i have ever worked with. I keep these locked up in my office each night, because i have found over the years that leaving them in my vehicle during extremely cold conditions, or in direct sun light on hot days damages or shortens the life. I have tried several other handhelds because of a cheaper price, but none lasted more than a few months in these harsh conditions. I use these for Groundwater, Title 27 Ponds, wastewaters, fresh water, concentrated waters, and prodcut waters. I can use this one handheld for all of my monitoring.
LOVE IT!

PROS
Easy Calibrations
Easy To Use
Exhibits Correct Properties
Quick Accurate Readings
CONS
NO Extreme hot cold temps

 

Categories : Image Ads, Product Updates